Table 2:

MR imaging parameters used in the assessment of the effects of MRI at 3T on the programmable valve

Pulse Sequence
T1-SET2-SET1-FSET2-FSEGRE, 3DGFRE, 3DGRE, MTCEPI
TR (ms)74130007005180201636283400
TE (ms)710091135410103
Flip angleN/AN/AN/AN/A25N/A25N/A
Field of view (cm)3030303012303030
Matrix size256 × 256256 × 256256 × 256256 × 256256 × 256256 × 256256 × 256256 × 256
Section Thickness (mm)10101010331010
Section gap (mm)11110.60.611
Imaging planeAxialAxialAxialAxialVolumeVolumeAxialAxial
Imaging time1:001:001:001:001:001:001:001:00
SAR, whole body (W/kg)2.00.62.31.10.80.31.50.5
  • Note:—T1-SE indicates T1-weighted spin-echo; T2-SE, T2-weighted spin-echo; T1-FSE, T1-weighted fast spin-echo; T2-FSE, T2-weighted fast spin-echo; GRE, gradient echo; FGRE, fast gradient echo; MTC, magnetization transfer contrast; EPI, echo-planar imaging; N/A, not applicable; SAR, specific absorption rate.