Table 2:

Comparison of the PD between flow measurements from QMRA and the flow probe across the normal and stenotic CCA as a function of stenosis

PositionPD (± SE) (%)No. of MeasurementsAverage Flow (mL/min)
No stenosis4.8 ± 133509
Stenosis11.4 ± 12786
    Proximal to stenosis9.6 ± 2882
    Distal to stenosis12.2 ± 21988