Table 2:

Definite vs no or questionable histologic damage to carotid arteries under various conditions of catheter testing in vivo

Condition TestedDefinite Damage (%)No/Questionable Damage (%)Odds Ratio95% CIP Value*
Arterial stasis (n = 24)4963.00.36–137.44
Normal flow (n = 70)1189
Current ≤300 mA (n = 58)010000–0.21.0001
Current > 300 mA (n = 36)2575
Activations ≤1 min (n = 52)6940.370.056–1.9.29
Activations > 1 min (n = 42)1486
Work ≤100 J (n = 68)3970.0820.008–0.5.0015
Work >100 J (n = 26)2773
Saline drip ≤2 mL/min (n = 39)15853.20.61–21.16
Saline drip > 2 mL/min (n = 55)595
≤5°C catheter tip coil temperature rise (n = 36)010000–0.36.0022
> 5°C temperature rise (n = 29)2476
MR and x-ray guidance (n = 45)11881.90.35–13.47
X-ray guidance only (n = 49)694
  • CI indicates confidence interval; J, Joules; mA, milliamperes.

  • Note:—Definite damage denotes a histologic score of 2; questionable damage, histologic score of 1; and no damage, histologic score of 0.

  • * P value = 2-tailed Fisher exact test.