Definite or questionable vs no histologic damage to carotid arteries under various conditions of catheter testing in vivo
Condition Tested | Definite/Questionable Damage (%) | No Damage (%) | Odds Ratio | 95% CI | P Value* |
---|---|---|---|---|---|
Arterial stasis (n = 24) | 21 | 79 | 0.87 | 0.25–3.5 | .77 |
Normal flow (n = 57) | 23 | 77 | |||
Current ≤300 mA (n = 51) | 14 | 86 | 0.31 | 0.092–1.020 | .034 |
Current > 300 mA (n = 36) | 31 | 69 | |||
Activations ≤1 minute (n = 52) | 8 | 92 | 0.17 | 0.037–0.61 | .003 |
Activations > 1 min (n = 42) | 33 | 67 | |||
Work ≤100 J (n = 68) | 13 | 87 | 0.29 | 0.018–1 | .037 |
Work > 100 J (n = 26) | 35 | 65 | |||
Saline drip ≤2 mL/min (n = 39) | 21 | 79 | 1.2 | 0.35–3.7 | .80 |
Saline drip >2 mL/min (n = 55) | 18 | 82 | |||
≤5°C catheter tip coil temperature rise (n = 36) | 8 | 92 | 0.17 | 0.028–0.80 | .013 |
> 5°C temperature rise (n = 29) | 34 | 68 | |||
MR and x-ray guidance (n = 45) | 16 | 84 | 0.64 | 0.19–2.0 | .44 |
X-ray guidance only (n = 49) | 22 | 78 |
↵* P value = 2-tailed Fisher exact test.