Table 3:

Definite or questionable vs no histologic damage to carotid arteries under various conditions of catheter testing in vivo

Condition TestedDefinite/Questionable Damage (%)No Damage (%)Odds Ratio95% CIP Value*
Arterial stasis (n = 24)21790.870.25–3.5.77
Normal flow (n = 57)2377
Current ≤300 mA (n = 51)14860.310.092–1.020.034
Current > 300 mA (n = 36)3169
Activations ≤1 minute (n = 52)8920.170.037–0.61.003
Activations > 1 min (n = 42)3367
Work ≤100 J (n = 68)13870.290.018–1.037
Work > 100 J (n = 26)3565
Saline drip ≤2 mL/min (n = 39)21791.20.35–3.7.80
Saline drip >2 mL/min (n = 55)1882
≤5°C catheter tip coil temperature rise (n = 36)8920.170.028–0.80.013
> 5°C temperature rise (n = 29)3468
MR and x-ray guidance (n = 45)16840.640.19–2.0.44
X-ray guidance only (n = 49)2278
  • * P value = 2-tailed Fisher exact test.